 APPLICATION NOTE 

Analysis on the influence of the sample’s quantity in pesticide extraction in an apple matrix using SiliaFast FaPEx cartridges

> back to SiliaFast FaPEx Applications Notes

Depending on the international standard, QuEChERS methods require using either 10 or 15 g samples. This compares to our SiliaFast FaPEx extraction cartridges, which require as little as 1 g of sample material. The present document shows a comparison on recovery values using a standard SiliaFast FaPEx cartridge with 1 g of sample and a custom-made cartridge with 10 g of sample.

« The results obtained show similar pesticide recoveries for both sample quantities used (1g & 10g) at two different concentrations (25 & 1000 ppb) using SiliaFast FaPEx. Neither the quantity of sample nor the concentration has a significant impact on the pesticide recovery. »