Follow SiliCycle on Facebook Follow SiliCycle on Twitter Follow SiliCycle on LinkedIn Watch our videos on Youtube

Halocetic Acids

> back to Suggested SPE Protocols

Matrix

Surface water

Analytes

Chloroacetic Acid, Bromoacetic Acid, Dichloroacetic Acid, Dalapon, Trichloroacetic Acid, Bromochloroacetic Acid, Bromodichloroacetic Acid, Dibromoacetic Acid, Chlorodibromoacetic Acid, Tribromoacetic Acid

SPE Cartridge

SiliaPrep SAX
500 mg / 6 mL
SPE-R66530B-06P

Sample Pretreatment before Loading

Add 0.5 mL of aqueous Ammonium Chloride to 50 mL of water sample. Adjust pH to 4.5 - 5.5 with Sulfuric Acid.

Conditionning

10 mL of Methanol

Equilibration

10 mL of Water

Loading

Pretreated sample

Washing

10 mL of Methanol

Elution

3 mL of 10 % Sulfuric Acid / Methanol

Further Treatment

Esterification and identification by GC/μECD.