Follow SiliCycle on Facebook Follow SiliCycle on Twitter Follow SiliCycle on LinkedIn Watch our videos on Youtube

Organophosphates

> back to Suggested SPE Protocols

Matrix

Surface water

Analytes

Tris (1-Chloro-2-Propyl)-Phosphate (TCPP), Tris (2-Chloroethyl)-Phosphate (TCEP), Tris (1,3-Dichloro-2-Propyl)-Phosphate (TDCP), Tri‑n‑Butylphosphate (TnBP), Tri-Isobutylphosphate (TiBP), Tris(2-Butoxyethyl)-Phosphate (TBEP)

SPE Cartridge

SiliaPrepX HLB
100 mg / 1 mL
SPE-P0002-01C

Sample Pretreatment before Loading

No sample pretreatment required.

Conditionning

1 mL of Methanol

Equilibration

1 mL of Methanol / Acetonitrile (50:50)

Loading

2 L of water sample
Dry the cartridge

Washing

No washing required

Elution

2 x 0.5 mL of Methanol / Acetonitrile (50:50)

Further Treatment

Quantification by GC/MS.

1 Item(s)

Set Descending Direction
per page

List  Grid 

  1. SiliaPrepX HLB Polymeric SPE Cartridges (SPE-P0002)

    SiliaPrepX HLB Polymeric SPE Cartridges, 40 µm, 110 Å (SPE-P0002)

    SiliaPrepX HLB is a wettable copolymer presenting a Hydrophobic-Lipophilic Balance (HLB) permitting a strong retention for neutral, acidic and basic compounds and a high stability in organic solvents.